Beer's Law & Spectrophotometry Simulator

Shine light through a virtual cuvette. Change the concentration, path length, and molar absorptivity to see how absorbance (A = εbc) and transmittance respond, then use a calibration curve to find an unknown.

Want the concepts first? Read the Beer–Lambert law lesson.

Illustrative model: the preset ε values, wavelengths and colors are teaching parameters, not measured reference data. The blue dye has no specified chemical identity. Assume a dilute solution, fixed wavelength, constant ε and a blank-corrected ideal instrument; real calibration depends on the species, solvent, temperature and instrument.

Illustrative wavelength: 525 nm. Changing ε explores the model; it does not measure a new spectrum or locate λmax.

Absorbance, A

0.480

Transmittance, %T

33.1%

A = εbc = 2400 × 1.0 × 2.00e-4 = 0.480

%T = 10−A × 100 = 33.1%

Light path

Sourceb = 1.0 cmDetector33%

Calibration curve

Five simulated standards calculated with the current ε and b. The slope of A vs. c equals εb = 2400 M⁻¹.

0.0000.662.63e-41.325.25e-4Concentration (M)Absorbance

c = A ÷ (εb) = 0.45 ÷ 2400 = 1.88e-4 M

AP FRQ traps

  • Water droplets or fingerprints on the cuvette scatter light → measured A is too high → calculated concentration is too high.
  • Rinsing the cuvette with water and leaving it wet dilutes the sample → A too low → concentration too low.
  • Measure at λmax: absorbance changes most with concentration there, giving the most sensitive calibration.
  • Absorbance has no units; ε has units of M⁻¹cm⁻¹ so that εbc is unitless.