SpectroscopyCourse level: AP Chemistry

Spectroscopy and the Beer-Lambert Law

What you'll be able to do: Use A = abc to relate absorbance to concentration, and read a calibration curve to find an unknown.

Introduction

A coloured solution absorbs light, and the more concentrated it is the more it absorbs. Spectroscopy turns that everyday observation into a precise measurement of concentration.

These are recommended, not required. You can start this lesson at any time.

Learning objectives

  • State and apply the Beer-Lambert law
  • Explain why measurements are taken at λ max
  • Construct and read a calibration curve
  • Identify sources of deviation from linearity

The law

A = abc, where A is absorbance, a is the molar absorptivity of the substance, b is the path length of the cuvette in centimetres, and c is the concentration in mol/L. Absorbance itself has no units, and molar absorptivity carries whatever units make the product dimensionless.

The common alternative symbol for molar absorptivity is ε, so A = εbc is the same relationship as A = abc. With b in cm and c in mol/L, ε has units L mol⁻¹ cm⁻¹. Absorbance is defined by A = −log₁₀(T), where transmittance T = I/I₀. T is a fraction, not a percent: 10% transmission means T = 0.10 and A = 1.00, while 1% means A = 2.00. Absorbance is linear in concentration within the applicable range, but percent transmission is not.

A = abc

With a and b held constant, absorbance is directly proportional to concentration. That single fact is what makes the whole technique work.

Choosing the wavelength

Every substance absorbs some wavelengths more than others. Measurements are taken at λ max, the wavelength of maximum absorbance, because the signal is largest there and small errors in wavelength setting have the least effect on the reading. A solution appears the complementary colour to the light it absorbs, so a solution that absorbs red light looks green.

At a wavelength where ε is larger, the calibration slope εb is steeper, so a small concentration change produces a larger absorbance change. Near a broad absorbance maximum the slope of absorbance versus wavelength is small, reducing sensitivity to small wavelength-setting differences. This is a reason to choose λ_max when no interfering species or instrument limitations dictate otherwise. The transmitted mixture of wavelengths gives the observed color; a single absorbed wavelength does not always map to one exact named complementary color.

Build and use a measured calibration line

Prepare a blank and several standards spanning the unknown's expected range, then measure them with the same wavelength, path length, solvent and relevant chemical conditions. Plot A vertically and concentration horizontally. The ideal blank-corrected relation passes through zero, but a measured best-fit line can have an intercept: A = mc + b₀. Use c = (A − b₀)/m rather than forcing the line through the origin without justification.

For A = (250 L/mol)c + 0.020 and Aunknown = 0.520, c = (0.520 − 0.020)/250 = 0.00200 M. Do not confuse the intercept b₀ with the cuvette path-length symbol b. Interpolate inside the standard range; if the unknown lies above it, dilute a measured aliquot, remeasure and multiply the inferred diluted concentration by Vfinal/V_aliquot to recover the original concentration. A 5.00 mL aliquot made to 50.00 mL has dilution factor ten.

Use the fitted calibration A = mc + b and its validated range. Subtract a justified nonzero intercept before dividing by slope.

When linearity fails

At high concentration the particles begin to interact and the line bends away from the axis, so very concentrated samples should be diluted into the linear range. Cloudiness scatters light and inflates the apparent absorbance, and a smudged or mismatched cuvette does the same.

Deviations can come from concentration-dependent interactions, changing chemical speciation, stray light or detector limitations. Turbidity removes light by scattering rather than only absorption, so it can inflate apparent absorbance and the inferred concentration. Fingerprints, bubbles, inconsistent cuvette orientation and a changed path length can also alter readings. Keep conditions matched and investigate nonlinearity rather than extending the calibration blindly.

Trace an error through the blank and slope

Blanking removes the cuvette and solvent background when performed as specified by the instrument. If readings are already blank corrected, subtracting the blank again would introduce an error. Use the same correction convention for standards and unknowns; a regression intercept may already represent residual background.

Suppose every standard is actually diluted to 80% of its labeled concentration while its x-axis label is left unchanged. At each labeled concentration the measured absorbance is too low, so the fitted slope is 80% of the true slope. For an otherwise correctly measured unknown, c_reported = A/slope is then too high by 1/0.80 = 1.25. The instrument is not seeing more analyte; the incorrect calibration makes a given absorbance appear to require a larger concentration. If instead only the unknown is diluted without correcting its dilution factor, its reported original concentration is too low. Name the affected sample before deciding the direction of error.

Match the calibration’s blank convention; do not subtract a background a second time from an already corrected reading.

Key ideas

Definition
Absorbance

A unitless measure of how much light a sample removes from the beam.

Definition
Molar absorptivity

A constant describing how strongly one particular substance absorbs at a given wavelength.

Definition
Lambda max

The wavelength of maximum absorbance, used for the most sensitive measurement.

Rule
Proportionality

With fixed a and b, doubling the concentration doubles the absorbance.

Equation
Beer-Lambert law

A = abc

  • A = absorbance, unitless
  • a = molar absorptivity
  • b = path length in cm
  • c = concentration in mol/L

Worked examples

Worked example 1

A solution has an absorbance of 0.420 in a 1.00 cm cuvette. The molar absorptivity is 2100 L/(mol*cm). Find the concentration.

Try it first: Rearrange the law for the unknown before substituting numbers.

    0 of 3 steps revealed.

    Worked example 2

    Standards of 0.0020 M and 0.0040 M give absorbances of 0.30 and 0.60. An unknown reads 0.45. Find its concentration.

    Try it first: Check whether the standards fall on a straight line through the origin.

      0 of 4 steps revealed.

      Worked example 3

      A blank reads 0.020. A sample reads 0.520. A blank-corrected 0.0100 M standard reads 0.250 at the same wavelength and path length. Find the sample concentration.

      Try it first: Remove the blank contribution first.

        0 of 3 steps revealed.

        Common mistakes

        Common mistakes and better checks
        MistakeWhy it's wrongCheck instead
        Treating absorbance as having units.Absorbance is a logarithmic ratio of light intensities, so it is dimensionless.Report absorbance as a bare number.
        Extrapolating a calibration line beyond the standards.Linearity is only demonstrated over the range that was measured.Dilute the unknown until it falls inside the calibrated range.
        Assuming the solution is the colour it absorbs.You see the transmitted light, which is the complementary colour.A green solution absorbs red light, not green.

        Scroll horizontally if needed to read all columns.

        After the lesson · Question bank

        Independent practice

        Work through new problems on your own. The lesson's quick checks and worked examples stay above if you need to review a step.

        35 topic questions across 3 difficulty levels.

        Loading practice questions…

        What you should now know

        Within the calibrated linear range at fixed wavelength and path length, absorbance is proportional to concentration. Use the measured calibration equation, including a supported intercept, and apply blank corrections consistently. If an aliquot is diluted, recover the original concentration with the dilution factor; scattering is not analyte absorption.

        • A = εbc connects absorbance, path length and concentration within its applicable range.
        • Absorbance is −log₁₀(transmittance) and has no units.
        • λ_max often gives high sensitivity and reduced wavelength-setting sensitivity.
        • Use c = (A − intercept)/slope for the stated fitted calibration.
        • Dilute into the calibrated range and account for dilution; investigate turbidity, speciation and instrument effects.

        Your next step

        Lesson: Not marked completed

        Marking a lesson completed records that you finished reading it. Practice establishes mastery separately. Saved on this device.

        Try a short set of different questions to check what you can do independently.

        Watch and Learn

        An optional video to support this lesson.

        Loading video resources…

        Supplementary learning resources

        More resources for this topic →

        External resources belong to their respective creators. Chem Help is not affiliated with or endorsed by these providers. Links open the original provider; their terms apply.

        Sources and further reading

        Source attribution for this lesson has not been verified. Missing source records do not establish originality or reuse permission.